you are expert in camera and lens, im developing microscope, and here is it
Camera - Lens - Sample (in glass)- Light
OBJ is moving lens and CAM is moving camera
so please try fix focus improve it and make it best focus getting fast and getting highest focus score
also make relation between zoom and moving OBJ focus and CAM to getting best focus fast here what i did and got best score focus
in zoom: 2.1
just made +1 mm of OBJ and changes in CAM and got best focus
in zoom: 6.6
just made + 0.9 mm of OBJ and +1 mm in CAM
in zoom: 7.5
just made OBJ + 1mm and CAM + 2.2mm
so take this refernce to make like equation or something to reach faster, so make the best algorithm to get highest focus possible in less time
critical: study case well and search for that to know everything and planning and building a good algorithm and return all updated files
I’ll treat this as an optics + control-algorithm problem, not just a code tweak. I’m going to map your current zoom/focus flow, derive a zoom→OBJ/CAM predictive starting model from your measurements, compare it with proven autofocus strategies, then update the four files so the search starts near the expected peak and refines quickly and safely.
The first important finding is already clear: your current “focus-only” button always searches OBJ in the negative direction first, then searches CAM, even though your real measurements say the optimum is consistently about +0.9 to +1.0 mm on OBJ from the zoom-calibrated position. That wastes several moves before it reaches the useful side of the focus curve.
A second code-level problem is more serious: after a zoom move, onZoomSequenceFinished() launches the old OBJ/CAM cell-count search, but the only callback that advances that search is commented out. So the zoom-triggered “autofocus” path can stall or behave inconsistently. I’m replacing that path with the focus-score engine and keeping the old cell search as a fallback utility instead of the main zoom-focus mechanism.
The algorithm I’m implementing is now: zoom calibration → predictive residual correction → direction-aware coarse bracket → adaptive robust sampling → quadratic sub-step peak estimate → stationary verification → learn the residual for that zoom. The prediction is only a seed; the measured focus score still decides the final position, so a bad model cannot silently “lock” focus at the predicted point.
I’ve tightened the fast path further: when the first move lands at the model-predicted correction and its score is clearly better, the search no longer wastes a second coarse probe. It immediately switches to a local fine bracket and quadratic peak verification; if the prediction is wrong, the existing bidirectional hill-climb fallback still takes over.
pythonimport numpy as np, math z=np.array([2.1,6.6,7.5]) obj=np.array([1.0,0.9,1.0]) coef=np.polyfit(z,obj,1) coef array([-0.00716846, 1.00537634])
pythonA=0.03758806; B=0.626557 def base(z): L=math.log(z) obj=((0.1760222*L-1.6729109)*L+5.1422488)*L+15.2942523 cam=(z-B)/A-obj return obj,cam for zz in [2.1,6.6,7.5]: o,c=base(zz) print(zz,o,c,o+c,(zz-B)/A)
pythonimport math, numpy as np, pandas as pd obj_pts=[(2.1,1.0),(6.6,0.9),(7.5,1.0)] cam_pts=[(6.6,1.0),(7.5,2.2)] def interp_obj(z): if z<=obj_pts[0][0]: return obj_pts[0][1] if z>=obj_pts[-1][0]: return obj_pts[-1][1] for (x1,y1),(x2,y2) in zip(obj_pts,obj_pts[1:]): if x1<=z<=x2: t=(z-x1)/(x2-x1) return y1+(y2-y1)*t def interp_cam(z): if z<=cam_pts[0][0]: a,b=cam_pts slope=(b[1]-a[1])/(b[0]-a[0]) v=max(0,a[1]+slope*(z-a[0])) return v, v>0.05 if z>=cam_pts[-1][0]: return cam_pts[-1][1],True a,b=cam_pts t=(z-a[0])/(b[0]-a[0]) return a[1]+(b[1]-a[1])*t,True rows=[] for z in [1.655,2.1,4,5.85,6,6.6,7,7.5,7.892]: od=interp_obj(z) cd,valid=interp_cam(z) rows.append((z,od,cd,valid)) pd.DataFrame(rows,columns=['z','objd','camd','camvalid']) z objd camd camvalid 0 1.655 1.000000 0.000000e+00 False 1 2.100 1.000000 0.000000e+00 False 2 4.000 0.957778 0.000000e+00 False 3 5.850 0.916667 2.220446e-16 False 4 6.000 0.913333 2.000000e-01 True 5 6.600 0.900000 1.000000e+00 True 6 7.000 0.944444 1.533333e+00 True 7 7.500 1.000000 2.200000e+00 True 8 7.892 1.000000 2.200000e+00 True
The updated search keeps the useful parts you already had—non-blocking motor handling, live peak observation, robust multi-frame scoring, and backlash-aware moves—but removes fixed-step assumptions where zoom gives us better information. I’m doing a final code-path/compile-sanity pass now, especially around startup completion, cancellation, motor timeouts, and fallback-to-best behavior so the speed changes don’t make the system brittle.
python# Let's calculate empirical formulas and predicted base/final positions for documentation. import math A=0.03758806 B=0.626557 def base(z): L=math.log(z) obj=((0.1760222*L-1.6729109)*L+5.1422488)*L+15.2942523 cam=(z-B)/A-obj return obj,cam for z in [2.1,6.6,7.5]: print(z,base(z))
I studied the optics, your existing search flow, and the three measured zoom/focus points, then rebuilt the zoom-triggered autofocus path around a predictive + measured closed loop.
Your existing zoom geometry already calculates a nominal OBJ position from a cubic in ln(zoom), then calculates CAM from the calibrated OBJ+CAM relationship. I kept that calibration intact rather than replacing it with only three measurements. (mainwindow.cpp)
After a zoom completed, the code was calling startObjSearch(). (mainwindow.cpp) That search does not optimize optical sharpness—it averages detected sperm/cell counts and chooses the motor position with the highest count. (mainwindow.cpp) On top of that, the motor callback that would advance that cell search is commented out. (mainwindow.cpp)
The actual sharpness autofocus also always started OBJ in the negative direction, with fixed 1 mm OBJ and 2 mm CAM coarse steps. (mainwindow.cpp) That conflicts directly with your measurements: at 2.1, 6.6, and 7.5 your useful OBJ correction is positive, and at 6.6/7.5 CAM is also positive.
I kept your existing geometric zoom model and added a focus residual:
For OBJ, the initial model interpolates your actual points:
So practically, OBJ focus correction is approximately +1 mm, but the code uses piecewise interpolation instead of pretending three measurements justify a complicated polynomial.
For CAM, you supplied two numeric calibration points:
which gives:
around the calibrated high-zoom region. I deliberately do not extrapolate that aggressively into low zoom. Below about 5.9× it falls back to measured autofocus until the system learns real data there.
With your existing zoom equation, the initial predictions are:
| Zoom | Existing OBJ base | Existing CAM base | Predicted first OBJ probe | Predicted first CAM probe |
|---|---|---|---|---|
| 2.1 | 18.260 mm | 20.939 mm | 19.260 mm | no guessed CAM value |
| 6.6 | 20.224 mm | 138.695 mm | 21.124 mm | 139.695 mm |
| 7.5 | 20.304 mm | 162.559 mm | 21.304 mm | 164.759 mm |
This also agrees qualitatively with imaging optics: focus couples object distance, lens position and image/sensor distance through the imaging equation, while magnification is related to image/object distance. For a real multi-element zoom lens, however, a simple thin-lens formula is not accurate enough by itself, so empirical calibration plus closed-loop sharpness verification is the safer approach. (www.edmundoptics.com www.rp-photonics.com)
The important change is that the equation is not allowed to decide the final focus. It only tells the algorithm where to look first.
After zoom, it measures a stationary baseline. OBJ then jumps in the direction and distance predicted for that zoom. If that first predicted point clearly improves focus, it immediately skips the unnecessary second coarse probe and enters fine search. If the prediction is wrong, the bidirectional hill-climb automatically takes over.
Fine steps now shrink with zoom, from about 0.10 → 0.06 mm for OBJ and 0.35 → 0.15 mm for CAM. That is appropriate because microscope depth of field generally becomes increasingly restrictive as magnification/NA increase. (www.microscopyu.com www.microscopyu.com)
Around the measured peak I added a 3-point concave quadratic fit. It estimates a sub-step maximum, physically moves to that estimated vertex, and measures it before accepting it. Recent microscopy autofocus research specifically reports that near-focus sharpness can be approximated by a negative quadratic and that curve fitting can improve precision beyond the physical scan step while reducing repeated motion. (pmc.ncbi.nlm.nih.gov)
I also kept robust multi-frame measurement but made it adaptive. Fine positions can collect up to 9 frames, but after 5 frames the algorithm checks median absolute deviation. If the score is already stable, it continues immediately; noisy positions still get the full burst. Motor settling is now distance-dependent as well.
This approach is also consistent with mature microscopy autofocus implementations: Micro-Manager supports derivative-free Brent optimization, coarse/fine search, and reduced camera ROI to cut autofocus time. (micro-manager.org micro-manager.org)
A successful zoom-triggered run saves:
zoom → best OBJ residual + best CAM residual + focus score
using QSettings.
So after you collect real focus results at, for example, 3×, 4×, 5×, 6× and 7×, those learned values replace/interpolate the initial three-point model automatically. Repeated measurements near the same zoom use a smoothed update instead of overwriting calibration from one potentially noisy run.
I also limited predictive corrections to ±2.5 mm OBJ and ±4 mm CAM. If calibration becomes bad, the ordinary image-score search still runs instead of allowing the predictor to command a huge movement.
One limitation remains: the actual sharpness metric is generated by DetectionThread::focusScoreUpdated. The supplied autofocus.cpp/.h explicitly says that class optimizes lighting/exposure, not sharpness. (autofocus.h) So I intentionally left those two files behaviorally unchanged. To improve the focus score itself—for example with a central ROI plus normalized Tenengrad/Laplacian-type metric—I would need detectionthread.cpp/.h.
I performed structural/static consistency checks on the edited sources and verified all added declarations/definitions and brace balance. A complete application compile isn't possible from this four-file snapshot because project headers such as detectionthread.h, hardware_control.h, sidepanel.h, analysisform.h, and generated ui_mainwindow.h were not included.
[Download the complete autofocus update bundle]
Updated mainwindow.cpp
[Updated mainwindow.h]
autofocus.cpp
[autofocus.h]
Algorithm, equations, and hardware test plan
[mainwindow.cpp diff]
mainwindow.h diff
The most valuable next engineering step is to run the included test protocol at roughly 2.1, 3, 4, 5, 6, 6.6, 7, and 7.5×. Once those real best positions accumulate, the predictor becomes a calibration map for your actual lens/sample/glass mechanics rather than a generic autofocus search.